Question
Why after installation does the sensor not operate correctly (the laser line flickers)?
Answer
Sensor Model | G2xxx |
Firmware Version | 3.x, 4.x |
SDK Version | n/a |
Ensure the system is properly grounded. Verify that the ground connection of the power supply and the shield of the Gocator power cable are connected to the same ground level.
Comments
0 comments