Sensor Model | G1xxx, G2xxx, G3xxx |
Firmware Version | All |
SDK Version | n/a |
Question
What is the resolution of Gocator's analog output?
Answer
The resolution of Gocator's analog output is fixed at 12 bits, or 4096 steps over a sensor's maximum current range.
The resolution does not change when the Current Range setting is changed in the Gocator web interface (Output page, in the Analog category). This means that if a smaller range is chosen, only a subset of the 4096 steps is available for output.
For legacy Gocator 2000 series sensors, the resolution is 4.88 µA.
For all other Gocator sensors, the resolution is 5.37 µA.