Sensor Model | G2xxx |
Firmware Version | 3.x, 4.x, 5.x, 6.x |
SDK Version | n/a |
Why don't I see live profile data in the Gocator web interface?
To see live profile data in the web interface, ensure that the trigger mode is configured properly and that the Gocator is running. The CPU and Speed indicators at the top of the web interface indicate if the Gocator is running.
For laser-based sensors, verify that the Laser Safety Signal is enabled and that the laser line is visible.
Make sure the target is within the measurement range of the sensor.
Only Gocator firmware version 3.x: When using a FireSync Master, select the correct device in the Connection menu of the Gocator.
Finally check, if laser safety adhesive tape on sensor laser diode has been removed.
Comments
0 comments
Please sign in to leave a comment.