Question
What is the unit of the timestamp in the Gocator CSV export file?
Answer
Sensor Model | All |
Firmware Version | 3.x, 4.x, 5.x, 6.x |
SDK Version | n/a |
Gocator sensors and Master network controllers use an internal time unit equal to 1/1.024 microseconds. This is the unit of the timestamp in the CSV file exported from the Gocator. However, the unit in the web interface is in microseconds.
Example:
If the sensor speed has been configured for 100 Hz operation, then the difference between two consecutive time stamps in the CSV file will be 10240. 10240/1.024 = 10000 microseconds = 0.01 seconds, f=100 Hz.