GoWebScan Calibration Process Not Completing
Why is my GoWebScan process not completing after scanning the calibration bar?
If the calibration process is stuck on "System is scanning. System will stop when calibration is complete." There are several factors to consider :
- In the settings.xml file: ensure Encoder Resolution and travel orientation are correctly set up.
- Ensure the encoder counts are updating in the Gocator Web UI when running the chain
- Ensure to run the calibration bar at least 6" past the laser line
- If running debug mode, it may take up to a minute to finish processing the calibration