   Rescuing Gocator firmware

Question 

My Gocator sensor won't boot. How do I rescue the firmware?

Answer 

A Gocator can become unbootable if an upgrade is stopped unexpectedly (for example, because of power failure or loss of connection). If this happens you must "rescue" the firmware.

To rescue the sensor's firmware:

  1. Access the LMI Device Rescue page on the sensor by adding "/rescue" to the IP address of the sensor (for example, http://192.168.1.10/rescue).
  2. Follow the instructions on the page.