   c+s 3xxx sensor replacement procedure

Question 

How to install a new or repaired sensor on an existing system?

Answer 

Mount and connect the sensors on the frame.

  1. Download the latest Client Installer and Firmware Release files from http://www.lmi3d.com/support/downloads.
  2. Delete any previous Client installation folders on Client PC.
  3. Install the downloaded Client software on the client PC.
  4. Start the FireSync kClient.exe application from the newly installed software.
  5. Select Connect item from the menu or click the Connect icon.
  6. Using the Basic connection option, enter the IP address of the server PC, then click OK.
  7. Once connected, in the device tree enter the correct serial numbers for the replaced sensors (those should be identified as missing).
  8. Next, select the Server node in the device tree. 
  9. From the Server menu, select Upgrade...
  10. Browse for the desired upgrade package (.dat file downloaded in step 1.), then click OK to begin the upgrade process.
  11. Wait for the upgrade process to complete.
  12. Power-cycle all system components (sensors, stations, master, slaves).
  13. After the client PC boots up, restart the FireSync kClient.exe application.
  14. Verify that the settings are correct and all sensors are listed in the device tree.
  15. Enable system functionality, report any issues.
  16. Calibrate the system.
  17. Disconnect the FireSync kClient.exe application.
  18. Now the system is in a ready state for your application to run.