   Gocator showing manual calibrated when I did not use the Calibration option

Question 

Why is the Gocator showing "Manually Calibrated" when I did not use the Calibration option?

Answer 

Sensor Model All
Firmware Version 3.x
SDK Version n/a

The resolution that defines the encoder tick/mm conversion is part of the calibration process. When this value is manually entered, the system will show the Manually Calibrated status. Using the Travel Calibration option, the Gocator will perform a travel and alignment calibration and define the resolution (mm/tick) value. The system will now show the Auto-Calibrated status.