   Analog output resolution

Question 

What is the resolution of Gocator's analog output?

Answer 

The resolution of Gocator's analog output is fixed at 12 bits, or 4096 steps over a sensor's maximum current range.

The resolution does not change when the Current Range setting is changed in the Gocator web interface (Output page, in the Analog category). This means that if a smaller range is chosen, only a subset of the 4096 steps is available for output.

For legacy Gocator 2000 series sensors, the resolution is 4.88 µA.